Affiliation:
1. University of California at Los Angeles Medical School
Abstract
The problems inherent in using frequency-specific stimuli in the sound field to determine threshold sensitivity are reviewed, including a discussion of some of the specific problems encountered when introducing pure tones, narrow bands of noise, and frequency-modulated (FM) tones. The results of two experiments are reported. In Experiment I, the relationship between pure tones and frequency-modulated tones is developed under earphones in an anechoic chamber, and in two sound-isolated auditory test rooms (not anechoic). Experiment I resulted in the development of a reference threshold sound pressure level for frequency-modulated signals in the sound field. In Experiment II the reference level was applied to a clinical test facility and evaluated with a group of hearing-impaired individuals. The results suggest that the sound-field reference levels accurately reflect monaural threshold under earphones, when the earphone is calibrated to the ANSI, 1969 standard, and the sound field is calibrated to the suggested standard.
Publisher
American Speech Language Hearing Association
Cited by
40 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献