Real-Ear-to-Coupler Difference: Physical and Perceptual Differences

Author:

Jorgensen Lindsey123,Barrett Rachel1,Jedlicka David34,Messersmith Jessica1,Pratt Sheila34

Affiliation:

1. Department of Communication Sciences and Disorders, University of South Dakota, Vermillion

2. VA Sioux Falls Healthcare System, SD

3. VA Pittsburgh Healthcare System, PA

4. Department of Communication Sciences and Disorders, University of Pittsburgh, PA

Abstract

Purpose: The real-ear-to-coupler difference (RECD) is a recommended measure for accurate hearing aid fittings, especially for pediatric populations. However, for adults, many clinicians question whether it is necessary. Method: Hearing aids were fit on two groups of 85 older adults seen at a Veterans Administration audiology clinic. One group was fit using RECD measurements, whereas the second group was fit with population-based average RECD values. The two groups had similar pure-tone hearing thresholds. Results: Like previous studies, there was little difference between the measured RECD for the right and left ears among the participants. Although the majority of the measured RECDs were within 1 SD of the mean, approximately 20% of those measured were outside of the normal range. It also was found that all participants produced lower (improved) Hearing Handicap Inventory for Elderly–Screening (HHIE-S) scores from pre- to postfitting, thus suggesting a reduction in self-perceived hearing handicap. Conclusions: Despite having similar prefitting HHIE-S scores, those participants who had their hearing aids fit using measured RECD values had lower postfitting scores than the group that was fit with average RECD values. Furthermore, there was a significant difference between the groups on several questions of the International Outcome Inventory–Hearing Aids, suggesting that there was higher satisfaction with the fittings based on the custom RECD rather than the fittings based on the average RECD. This study demonstrated that, in addition to performing verification using real-ear measurements, accurate conversion of dB HL to dB SPL using personalized RECD likely improved hearing aid satisfaction.

Publisher

American Speech Language Hearing Association

Subject

Speech and Hearing

Reference27 articles.

1. Manufacturer Predicted And Measured REAR Values In Adult Hearing Aid Fitting: Accuracy And Clinical Usefulness

2. Guidelines for the audiological management of adult hearing impairment;American Academy of Audiology;Audiology Today,2006

3. American National Standards Institute. (2013). Methods of measurement of real-ear performance characteristics of hearing aids (S3.46–2013).

4. Middle ear resonance in infants: Age effects

5. Real-Ear-to-Coupler Difference Predictions as a Function of Age for Two Coupling Procedures