Abstract
Purpose
Studying normal or disordered motor control requires accurate motion tracking of the effectors (e.g., orofacial structures). The cost of electromagnetic, optoelectronic, and ultrasound systems is prohibitive for many laboratories and limits clinical applications. For external movements (lips, jaw), video-based systems may be a viable alternative, provided that they offer high temporal resolution and submillimeter accuracy.
Method
The authors examined the accuracy and precision of 2-D and 3-D data recorded with a system that combines consumer-grade digital cameras capturing 60, 120, or 240 frames per second (fps), retro-reflective markers, commercially available computer software (APAS, Ariel Dynamics), and a custom calibration device.
Results
Overall root-mean-square error (RMSE) across tests was 0.15 mm for static tracking and 0.26 mm for dynamic tracking, with corresponding precision (
SD
) values of 0.11 and 0.19 mm, respectively. The effect of frame rate varied across conditions, but, generally, accuracy was reduced at 240 fps. The effect of marker size (3- vs. 6-mm diameter) was negligible at all frame rates for both 2-D and 3-D data.
Conclusion
Motion tracking with consumer-grade digital cameras and the APAS software can achieve submillimeter accuracy at frame rates that are appropriate for kinematic analyses of lip/jaw movements for both research and clinical purposes.
Publisher
American Speech Language Hearing Association
Subject
Speech and Hearing,Linguistics and Language,Language and Linguistics
Reference31 articles.
1. Ariel Dynamics. (2012). Product catalog. Retrieved from www.arielnet.com/cloud/Catalogs/Show/Online
2. Accuracy of the NDI Wave Speech Research System
3. BTS Bioengineering. (2012). BTS Smart DX: High-precision optoelectronic system for the biomechanical motion analysis. Retrieved from www.btsbioengineering.com/products/kinematics/bts-smart-dx/
4. Carstens Medizinelektronik. (2012). High precision electromagnetic articulograph AG501. Retrieved from www.articulograph.de
5. The feasibility of a kinematic measure of lip closure during meaningful speech
Cited by
20 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献