Abstract
Abstract
This study focuses on the experimental demonstration of the definition of kelvin using a triple point of water (TPW) cell and estimates the measurement uncertainty. The first part of this article is to construct the measurement setup and perform the measurements. The second stage is related with the calculation of uncertainty according to guides to the expression of uncertainty in measurement. The experiment in this study can be a practical and convenient method to show the unit as the fraction 1/273.16 of the thermodynamic temperature of the TPW, 273.16 K.
Subject
General Physics and Astronomy,Education
Reference13 articles.
1. Good practice guide no. 125;Rusby,2016
2. Guide to the realization of the ITS-90 triple point of water;Peruzzi;Bureau International des Poids et Mesures,2017
3. The kelvin redefined;Machin;Meas. Sci. Technol.,2018
4. Redefinition of the kilogram, ampere, kelvin and mole: a proposed approach to implementing CIPM recommendation 1 (CI-2005);Mills;Metrologia,2006