Development and validation of a novel fibre-optic respiratory rate sensor (FiRRS) integrated in oxygen delivery devices

Author:

Sinha R,Hernandez Francisco U,He CORCID,Korposh SORCID,Correia RORCID,Norris A M,Liu C,Hayes-Gill B R,Morgan S PORCID

Abstract

Abstract Respiratory rate (RR) monitoring provides crucial information on the overall health condition of patients and a reliable, low cost RR monitor for normal hospital inpatient or home use would be of significant benefit. The proposed system measures light reflection from a Fibre Bragg Grating (FBG) located near, and the total reflection spectrum from a humidity sensing film deposited at, the tip of an optical fibre. Every breath causes a shift in the wavelength reflected from the FBG and intensity change in the overall reflection spectrum. The accuracy of different techniques is investigated in a two-part study with 15 healthy volunteers. In part 1, the participants’ respiration rate followed a handheld mobile application at 5, 12 and 30 breaths per minute with simultaneous measurement using the optical fibre system, thoracic impedance pneumography (TIP) and capnometry device (where possible). Two types of medical face masks and a nasal cannula with oxygen delivery rates were investigated. In part 2, participants wore an anaesthetic face mask and breathed at normal and low tidal volumes to evaluate whether low tidal volumes could be detected. The most accurate measurement of RR was through monitoring the Bragg wavelength shift (mean accuracy = 88.1%), followed by the intensity change at the Bragg wavelength (mean accuracy = 78.9%), capnometry (mean accuracy = 77.8%), area under the overall spectrum (mean accuracy = 65.4%) and TIP (mean accuracy = 43.1%). The Fibre-optic Respiratory Rate Sensor system (FiRRS) can differentiate between normal and low tidal volumes (p-value < 0.05) and demonstrated higher accuracy than capnometry measurement of end-tidal carbon dioxide in exhaled air. These latter two monitors measured RR more accurately than TIP. A comparable accuracy in the measurement of RR was obtained when the FiRRS was implemented in nasal cannula and face masks.

Funder

Engineering and Physical Sciences Research Council

Publisher

IOP Publishing

Subject

Surfaces, Coatings and Films,Acoustics and Ultrasonics,Condensed Matter Physics,Electronic, Optical and Magnetic Materials

Cited by 13 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3