Abstract
Abstract
Magnetic force microscopy (MFM) has become one of the most important instruments for characterizing magnetic materials with nanoscale spatial resolution. When analyzing magnetic particles by MFM, calibration of the magnetic tips using reference magnetic nanoparticles is a prerequisite due to similar orientation and dimension of the yielded magnetic fields. However, in such a calibration process, errors caused by extra electrostatic interactions will significantly affect the output results. In this work, we evaluate the magnetic moment and dipole radius of the MFM tip on Fe3O4 nanoparticles by considering the associated electrostatic force. The coupling of electrostatic contribution on the measured MFM phase is eliminated by combining MFM and Kelvin probe force microscopy together with theoretical modeling. Numerical simulations and experiments on nickel nanoparticles demonstrate the effectiveness of decoupling. Results show that the calibrated MFM tip can enable a more accurate analysis of micro-and-nano magnetism. In addition, a fast and easy calibration method by using bimodal MFM is discussed, in which the acquisition of multiple phase shifts at different lift heights is not required.
Funder
National Natural Science Foundation of China
Strategic Priority Research Program (C) of the CAS
Subject
Electrical and Electronic Engineering,Mechanical Engineering,Mechanics of Materials,General Materials Science,General Chemistry,Bioengineering
Cited by