Author:
Kossel L,Pfotenhauer J,Harter R
Abstract
Abstract
Cryo-ablation is a breast cancer treatment method that utilizes a small probe in precise locations within the body to freeze and destroy unwanted cancer tissue. Recently, there has been a growing interest in combining cryo-ablation with magnetic resonance imaging (MRI). The challenge of combining these two technologies is that MRI devices require the absence of metals, and traditional heat exchangers used in cryoprobes are typically composed of thermally conductive metals that tend to disrupt the image produced by an MRI, impairing its functionality. Subsequently, it becomes of interest to develop a heat exchanger composed of thermally conductive MRI-compatible materials such as zirconium. This report presents the results of a thermal modelling effort to characterize and design a non-metallic Joule-Thomson cryoprobe for cryo-ablation. The model is comprised of a Joule-Thomson valve, as well as a discretized recuperative heat exchanger that includes the effects of axial conduction, pressure drop, two-phase and single-phase convection correlations, and fluid properties for single-components and mixtures. The device operates at 170 K with a heat load of 10 W, which is a viable temperature and heat load for cryo-ablation.
Reference13 articles.
1. Joule thompson cryocoolers and cryoablation;Skye,2018
2. Mixture optimization for mixed gas joule-thomson cycle;Detlor;IOP Conf. Series: Materials Science and Engineering,2017