Author:
Jain Vinamra,Dommeti Vamsi Krishna,Nutu Emil,Merdji Ali,Biswas Jayanta Kumar,Roy Sandipan
Abstract
Abstract
Implants have created a great impact in the bio medical industry. There are different kinds of dental implants which are existing in the market. Every implant has a unique feature and withstands for different time period. This study on the dental implant creates a new model which includes the porous and tapered angle. Titanium alloy was used for making the implant and supporting bone tissue was considered. Computed tomography data were used to generate the mandible model. In the analysis both static and dynamic loading condition were considered. Three dimensional FEM was performed on the implant to find out the mechanical response for this specific type of implant. The von-Misses stress were calculated for implants and it was observed that safe when compared with yield strength of implant and abundant. In analysis it was also varied bone quality for cortical and cancellous bone and the effect of stress produced can be predated before the implantation done for the patients.
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献