Predictive quality assurance for linear accelerator target failure using statistical process control

Author:

Li JingdongORCID,Wang Dongxu,Chan Maria

Abstract

Abstract The performance of a linear accelerator (Linac) depends on the integrity of its x-ray target. The sudden failure of its target not only breaks down the Linac but also could contribute significant disruptions to patient care. This work is to develop a predicative quality assurance (QA) method using Statistical Process Control (SPC) and AutoRegressive Integrated Moving Average (ARIMA) modeling to identify the risk of target failure before it occurs. In the past years, we observed two incidents of target failure among our Linacs. Retrospectively, we collected past daily QA data (from both open fields and enhanced dynamic wedge (EDW) measurements) and analyzed its historical trend using methods of SPC and ARIMA. SPC is a technique that monitors process performance based on statistical analysis. ARIMA is a time-series forecasting algorithm that can be used to estimate future values based on its past pattern. Both have been evaluated for predictive QA in radiotherapy. Application of SPC on open beam QA data would not yield an early warning signal to the pending target failures. However, when the same SPC methodology applies to EDW measurements, the control limits were breached a couple of weeks before the target failed. EDW mechanism introduces nonuniform magnification factors over its wedge-directed beam profiles and is responsible for the sensitivity of its profile to changing beam properties induced by a degrading target. Further extension of the warning period may be possible by using ARIMA modeling. Predicative QA for EDW daily data using SPC and ARIMA methods may provide an early QA warning to incoming Linac target failure.

Funder

National Institutes of Health

Publisher

IOP Publishing

Subject

General Nursing

Reference23 articles.

1. Task Group 142 report: quality assurance of medical accelerators;Klein;Med. Phys.,2009

2. Determination of absorbed dose in a patient irradiated by beams of x- or gamma-rays in radiotherapy procedures,1976

3. The report of Task Group 100 of the AAPM: application of risk analysis methods to radiation therapy quality management;Huq;Med. Phys.,2016

4. AAPM task group 198 report: an implementation guide for TG 142 quality assurance of medical accelerators;Hanley;Med. Phys.,2021

5. Quality in radiation oncology;Pawlicki;Med. Phys.,2007