Phantom assessment of three-dimensional geometric distortion of a dedicated wide-bore MR-simulator for radiotherapy

Author:

Law Max W KORCID,Yuan JingORCID,Wong Oilei O LORCID,Ding Abby Y,Zhou YihangORCID,Cheung Kin Y,Yu Siu K

Abstract

Abstract This study evaluated the machine-dependent three-dimensional geometric distortion images acquired from a 1.5T 700 mm-wide bore MR-simulator based on a large geometric accuracy phantom. With the consideration of radiation therapy (RT) application requirements, every sequence was examined in various combinations of acquisition-orientations and receiver-bandwidths with console-integrated distortion correction enabled. Distortion was repeatedly measured over a six-month period. The distortion measured from the images acquired at the beginning of this period was employed to retrospectively correct the distortion in the subsequent acquisitions. Geometric distortion was analyzed within the largest field-of-view allowed. Six sequences were examined for comprehensive distortion analysis—VIBE, SPACE, TSE, FLASH, BLADE and PETRA. Based on optimal acquisition parameters, their diameter-sphere-volumes (DSVs) of CT-comparable geometric fidelity (where 1 mm distortion was allowed) were 333.6 mm, 315.1 mm, 316.0 mm, 318.9 mm, 306.2 mm and 314.5 mm respectively. This was a significant increase from 254.0 mm, 245.5 mm, 228.9 mm, 256.6 mm, 230.8 mm and 254.2 mm DSVs respectively, when images were acquired using un-optimized parameters. The longitudinal stability of geometric distortion and the efficacy of retrospective correction of console-corrected images, based on prior distortion measurements, were inspected using VIBE and SPACE. The retrospectively corrected images achieved over 500 mm DSVs with 1 mm distortion allowed. The median distortion was below 1 mm after retrospective correction, proving that obtaining prior distortion map for subsequent retrospective distortion correction is beneficial. The systematic evaluation of distortion using various combinations of sequence-type, acquisition-orientation and receiver-bandwidth in a six-month time span would be a valuable guideline for optimizing sequence for various RT applications.

Publisher

IOP Publishing

Subject

General Nursing

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3