Calibration system correction factor for measuring the reference air kerma rate (RAKR) applied to high dose rate 192Ir sources (HDR)

Author:

da Silva Douglas BatistaORCID,da Silva Ademir XavierORCID,Alves Carlos Frederico Estrada,Magalhaes Luis Alexandre GonçalvesORCID

Abstract

Abstract The aim of this study was to use computer simulation to analyze the impact of the aluminum fixing support on the Reference Air Kerma (RAK), a physical quantity obtained in a calibration system that was experimentally developed in the Laboratory of Radiological Sciences of the University of the State of Rio de Janeiro (LCR-UERJ). Correction factors due to scattered radiation and the geometry of the 192Ir sources were also sought to be determined. The computational simulation was validated by comparing some parameters of the experimental results with the computational results. These parameters were: verification of the inverse square law of distance, determination of (RAKR), analysis of the source spectrum with and without encapsulation, and the sensitivity curve of the Sourcecheck 4PI ionization chamber response, as a function of the distance from the source along the axial axis, using the microSelectron-v2 (mSv2) and GammaMedplus (GMp) sources. Kerma was determined by activity in the Reference air, with calculated values of 1.725 × 10−3 U. Bq−1 and 1.710 × 10−3 U. Bq−1 for the ionization chamber NE 2571 and TN 30001, respectively. The expanded uncertainty for these values was 0.932% and 0.919%, respectively, for a coverage factor (k = 2). The correction factor due to the influence of the aluminum fixing support for measurements at 1 cm and 10 cm from the source was 0.978 and 0.969, respectively. The geometric correction factor of the sources was ksg = 1.005 with an expanded uncertainty of 0.7% for a coverage factor (k = 2). This value has a difference of approximately 0.2% compared to the experimental values.

Publisher

IOP Publishing

Reference18 articles.

1. Air kerma standard for calibration of well-type chambers in Brazil using 192Ir HDR sources and its traceability;Di Prinzio;Med. Phys.,2009

2. Calibration of 1 9 2Ir highdoserate afterloading systems;Goetsch;Med. Phis.,1991

3. Comparison of air-kerma strength determinations for HDR 192Ir sources;Brian;Med. Phys.,2011

4. Desenvolvimento e Implantação da Calibração de Câmaras de Ionização Tipo Poço com Fonte de 192Ir de Alta Taxa de Dose no Brasil;Di Prinzio,2009