Author:
Hato Tsunehiro,Tsukamoto Akira,Adachi Seiji,Tanabe Keiichi
Abstract
Abstract
We developed a hybrid cooling system consisting of a liquid-nitrogen Dewar and a cryocooler with the aim of cooling HTS-SQUIDs fast at operation sites, extending cooling period time, and avoiding noise increase caused by the cryocooler. Liquid nitrogen is evaporated mainly by thermal inflow into the Dewar. Thus, we tried reducing the thermal inflow into a glass Dewar by cooling its inner surface using a small Stirling cryocooler with a cooling capacity of 16 W at 77 K, and examined the cooling period and the operation procedure. We successfully kept 0.6 liter of liquid nitrogen for one week with 0.2 liter reduction. It was also indicated that long time and low noise operation of HTS-SQUIDs would be possible in the hybrid cooling system by temporarily stopping the cryocooler during measurements.
Subject
General Physics and Astronomy
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献