Abstract
Abstract
A sliding-window (SW) methodology for VMAT dose calculation was developed. For any two adjacent VMAT control points (CPs) n and n+1, the dose distribution was approximated by a 2-CP SW IMRT beam with the starting MLC positions at CP n and ending MLC positions at CP n+1, with the gantry angle fixed in the middle of the two VMAT CPs. Therefore, for any VMAT beam with N CPs, the dose is calculated with N-1 SW beams. VMAT plans were generated for ten patients in Pinnacle using 4° gantry spacing. For each patient, the VMAT plan was converted to a SW IMRT plan and dose was re-calculated. Another VMAT plan, with 1° gantry spacing, was created by interpolating the original VMAT beam. The original plans were delivered on an Elekta Versa HD and measured with Mapcheck2 using an in-house developed subarc method. For both the isodose distribution and DVH, there were significant differences between the original VMAT plan and either the SW or the interpolated plan. However, they were indistinguishable between the SW and interpolated plans. The average passing rate between the original VMAT plan and measurements was 84%. For both the interpolated and SW plans, the average passing rate was 96%. We conclude that the proposed SW approach improves VMAT dose calculation accuracy without increase in dose calculation time.