Determination of composite field correction factor kpcsr,msr for Tomotherapy

Author:

Fitriandini A,Wibowo W E,Pawiro S A

Abstract

Abstract The purpose of this study is to determine the correction factor of the composite field (kpcsr,msr ) of Tomotherapy for head and neck cases and to investigate the influence parameters of the pcsr field. This study employed Exradin A1SL and A 16 as the investigated chambers, and EBT3 as the reference dosimeter. The pcsr fields were created based on cylindrical water solid phantom. The results of this study show that the kpcsr,msr values for both detectors increased with increasing field width (jaw), decreasing complexity level, and increasing pitch value. The yielded results indicate several factors affecting the kpcsr,msr values: target homogeneity, volume averaging effect, thread effect, and the difference of general recombination at the different field width. These results represent that kpcsr,msr is specific and depends on the parameters of the composite field. The kpcsr,msr values for head and neck cases vary from 0.978-1.028. This variation implies that the determination of pcsr field should be according to the parameters used in clinical conditions. Due to the different position of the chamber and the film, the target should have excellent dose homogeneity to avoid the different response of the detectors.

Publisher

IOP Publishing

Subject

General Physics and Astronomy

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3