Grebenikova N M,Davydov V V
Abstract
Abstract
The main factors responsible for measurement errors in the determining of the refractive index of the flowing fluid are considered. It was found, that the most significant effect on the measurement error have changes in optical density and temperature of the investigated flowing fluid, scattering of radiation in cuvette transducer and wedge shape of the cuvette glass. A new design of differential refractometer has been developed to study the influence of the optical density values of the flowing fluid on the measurement errors. The results of experimental studies are presented.