Author:
Al-Khasawneh K.,Borris E.,Bruückner B.,Eberhardt K.,Erbacher P.,Fiebiger S.,Gernhäauser R.,Göobel K.,Heftrich T.,Kisselbach T.,Kurtulgil D.,Langer C.,Reich M.,Reifarth R.,Renisch D.,Thomas B.,Volknandt M.,Weigand M.
Abstract
Abstract
A new detector setup (NICE-detector) based on an organic plastic scintillator is presented. It will be used during experiments measuring neutron-induced reactions with a charged particle in the exit channel. The proposed design was tested at the Goethe University Frankfurt. One of the test cases was the capture cross-section of 209Bi at different astrophysically important energies, including thermal capture cross-section and resonance integral. This research presents the performance of the detector setup, as well as preliminary results for the calculated cross-sections. The preliminary results demonstrate that the newly developed NICE-detector can be used to determine capture cross-sections with sufficient accuracy, and might be adopted in the future as an alternative for charged particle measurements in different nuclear and astrophysical applications.