Author:
Hapiddin Asep,Yulita Ika P.,Boynawan A.M.,Ratnaningsih ,Agmal Swivano,Novyanto Okasatria
Abstract
Abstract
In the metrology area, typical stabilized lasers are used as length primary standard. KIM-1, i.e. the iodine stabilized He-Ne laser in National Measurement Standards Laboratory-National Standardization Agency of Indonesia (SNSU-BSN) has been traceable to SI through CCL-K11 inter-laboratory comparison. The result was suitable for KIM-1 to be used as the length primary standard in SNSU-BSN. Beat frequency measurement has been applied in the optical frequency and wavelength calibration system for stabilized He-Ne laser 633 nm using KIM-1 as a reference standard. In the calibration replica, a dual-frequency mode (Agilent 5519B) took a role as DUT laser, which emits a pair of beams with a central wavelength of 632.991 354 nm in the vacuum with frequency difference 3.4 to 4.0 MHz and ±0.02 ppm stability for a typical lifetime. As a measurement result, the beat frequencies of the 1st and 2nd polarized beam of Agilent 5519B against KIM-1 are (121.33±0.06) MHz and (118.59±0.06) MHz.
Subject
General Physics and Astronomy
Reference10 articles.
1. Practical realization of the definition of the metre, including recommended radiations of other optical frequency standards (2003);Quinn;Metrologia,2003
2. Report on on-going CCL Key Comparison for the year 2014 Comparison of optical frequency and wavelength standards Final;Michael;Metrol. Tech. Suppl.,2017
3. Absolute frequency measurement of the iodine-stabilized He-Ne laser at 633 nm;Yoon;Appl. Phys. B,2001