Beat frequency measurement of the stabilized He-Ne laser 633 nm calibration in SNSU-BSN

Author:

Hapiddin Asep,Yulita Ika P.,Boynawan A.M.,Ratnaningsih ,Agmal Swivano,Novyanto Okasatria

Abstract

Abstract In the metrology area, typical stabilized lasers are used as length primary standard. KIM-1, i.e. the iodine stabilized He-Ne laser in National Measurement Standards Laboratory-National Standardization Agency of Indonesia (SNSU-BSN) has been traceable to SI through CCL-K11 inter-laboratory comparison. The result was suitable for KIM-1 to be used as the length primary standard in SNSU-BSN. Beat frequency measurement has been applied in the optical frequency and wavelength calibration system for stabilized He-Ne laser 633 nm using KIM-1 as a reference standard. In the calibration replica, a dual-frequency mode (Agilent 5519B) took a role as DUT laser, which emits a pair of beams with a central wavelength of 632.991 354 nm in the vacuum with frequency difference 3.4 to 4.0 MHz and ±0.02 ppm stability for a typical lifetime. As a measurement result, the beat frequencies of the 1st and 2nd polarized beam of Agilent 5519B against KIM-1 are (121.33±0.06) MHz and (118.59±0.06) MHz.

Publisher

IOP Publishing

Subject

General Physics and Astronomy

Reference10 articles.

1. Practical realization of the definition of the metre, including recommended radiations of other optical frequency standards (2003);Quinn;Metrologia,2003

2. Report on on-going CCL Key Comparison for the year 2014 Comparison of optical frequency and wavelength standards Final;Michael;Metrol. Tech. Suppl.,2017

3. Absolute frequency measurement of the iodine-stabilized He-Ne laser at 633 nm;Yoon;Appl. Phys. B,2001

Cited by 5 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Photon energy and photon behavior discussions;Energy Reports;2022-05

2. Calculation of the Photon Speed and Photon Energy Discussions;Springer Proceedings in Physics;2022

3. The development software used for surface plate calibration in national standardization agency of Indonesia;PROCEEDING OF THE 1ST INTERNATIONAL CONFERENCE ON STANDARDIZATION AND METROLOGY (ICONSTAM) 2021;2022

4. Traceability of the optical wavelength 633 nm calibration at SNSU-BSN;THE INTERNATIONAL CONFERENCE ON ADVANCED MATERIAL AND TECHNOLOGY (ICAMT) 2021;2022

5. Study in determining of junction point value for shifting measurement method on line scale calibration in SNSU BSN;Journal of Physics: Conference Series;2021-02-01

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3