Author:
Wijaya Nur Hudha,Budimansyah ,Sukwono Djoko
Abstract
Abstract
High voltage (kV) on an X-ray machine is the voltage produced by the High Tension Transformer (HTT) which will give a potential difference to the X-ray machine’s tube. Regional General Hospitals (RSUD) in Indonesia still widely use X-ray machines with analog systems as kV selection. This will result in the radiographer being exposed to radiation continuously when running the X-ray machine because they have to operate the tool directly. The many uses of cables on analog systems are still less efficient, therefore a wireless X-ray machine controller is designed that can reduce radiation exposure to radiographers, and reduce the use of cables on X-ray machines. This study uses Personal Computer (PC) as a place to set parameter values, namely kV. Serial communication using Bluetooth HC-05, minimum system uses ATmega 328p IC and data processing on PC using Delphi application. Measurements on the device use a multimeter to determine the input voltage for HTT. The measurement results of the parameters 60 kV, 65 kV, 70 kV, 75 kV, and 80 kV obtained the smallest correction value is 0.15 kV on the 75 kV parameter and the largest is 1.66 kV on the 65 kV parameter. The correction value is still within the tolerance value range of ± 1kV. In the measurement of distance the farthest results obtained by 6 meters with a barrier wall thickness of 9.5 cm.
Subject
General Physics and Astronomy
Reference20 articles.
1. Perekayasaan Prototip Pesawat Sinar-X Berbasis Mikrokontroler;Suyatno,2010
2. Aplikasi radiasi sinar-x di bidang kedokteran untuk menunjang kesehatan masyarakat;Suyatno;SDM Teknologi Nuklir,2008
3. Radiasi Di Kawasan Batan Yogyakarta;Rikasjono,2008
4. Geometry calibration between X-ray source and detector for tomosynthesis with a portable X-ray system;Kohei Sato;Int. J. Comput. Assist. Radiol. Surg.,2017
5. Pembuatan sistem pengendali parameter tegangan, arus dan pew aktu pada pesawat sinar-x;Dan,2013
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. An improved performance of x-ray machine prototype with wireless ma control system based on microcontroller;AIP Conference Proceedings;2024
2. Baby scales with data transfer and database storage using Android;PROCEEDINGS OF THE 9TH INTERNATIONAL SYMPOSIUM ON INNOVATIVE BIOPRODUCTION INDONESIA ON BIOTECHNOLOGY AND BIOENGINEERING 2022: Strengthening Bioeconomy through Applied Biotechnology, Bioengineering, and Biodiversity;2023
3. Simple blood bag storage using peltier;THE 2ND INTERNATIONAL CONFERENCE OF SCIENCE AND INFORMATION TECHNOLOGY IN SMART ADMINISTRATION (ICSINTESA 2021);2022
4. X-Ray machine control with wireless based on mA parameters;IOP Conference Series: Materials Science and Engineering;2021-02-01
5. Decubitus pump health device implementation with pressure sensor based on Atmega 328;IOP Conference Series: Materials Science and Engineering;2021-02-01