Author:
Ma Lishuang,Sen Qian,Wu Qi,Zhu Yao,Yan Min,Peng Shuo,Wang Zhile
Abstract
This manuscript aims to use two ultra-fast photomultiplier tubes (FPMT) coupled with Lu1.8Y2SiO5:Ce (LYSO) crystals to perform a coincidence time resolution (CTR) test applied to Time of Flight-Positron Emission Tomography (TOF-PET). The FPMT used in this work is a kind of Micro Channel Plate-PMT(MCP-PMT) with rise time of 100ps, TTS of 46ps working in single photon mode. The scintillation light waveform of the LYSO crystal appears “separated” when it is detected by the FPMT. At the same time, it is found that the setting of the timing threshold has a great impact on the results of the time resolution. The waveform processing algorithm is optimized by interpolation, and the acquired waveform is processed by the leading edge timing method. The whole test is carried out under the radiation background of 22Na, the CTR (sigma) measured by LYSO crystal coupled with FPMT is 40ps.
Subject
General Physics and Astronomy
Reference7 articles.
1. Evaluation of Various Scintillator Materials in Radiation Detector Design for Positron Emission Tomography (PET);Xie;Crystals,2020
2. Comparative timing measurements of LYSO and LFS-3 to achieve the best time resolution for TOF-PET;Doroud;NIMA,2015
3. Characterization of three LYSO crystal batches. Nuclear Instruments and Methods;Yang;Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment,2015
4. Investigation of the time performance of a LYSO array for TOF-PET;Jun-Hui,2015
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献