The average photon flux is one of the main parameters of any photon source. At the European XFEL (EuXFEL), this is continuously monitored in the tunnels by X-ray Gas Monitors (XGMs) [1-3]. However, to measure the absolute value of the X-ray flux at the sample location in the scientific end-stations, in general a smaller device is required since the XGMs and their associated differential pumping stages are very space demanding. Therefore, a miniature solid-state room-temperature calorimeter based on a design by AIST, Japan [4-8] was commissioned with beam at the Spectroscopy and Coherent Scattering (SCS) instrument where it could be located directly downstream of an XGM. This allowed for validation of the gas-based measurements with a solid-state physics method with an independent absolute calibration. The calorimeter was then applied at the High Energy Density (HED) instrument to characterize the beamline transmission and to calibrate a commercial laser power meter (LPM) [9] as a secondary relative monitor for X-ray measurements. This contribution describes the technical parameters and capabilities of the calorimeter and LPM and presents the measurements taken at SCS and HED.