Author:
Zhao Xia,Wang Ying,Wu Jian,Jia Xin,Shi Qufei
Abstract
Abstract
The dry block temperature calibrator uses air as the calibration medium with small volume and light weight, and therefore is widely applied in field calibration work. With the improvement and vigorous development of the technology of domestic instruments, the technology of domestic dry body furnace is also developing rapidly. Whether the performance of domestic dry block temperature calibrator is better than the foreign instruments, this question is not very clear yet. Therefore, this study focuses on the comparative analysis of the performance parameters of the two dry-type temperature calibrators of Foreign dry furnace-RTC158 and domestic dry furnace-Const660, such as temperature deviation, temperature volatility, axial deviation, and radial deviation. The results reveal that the comprehensive performance of domestic instrument-Const660 is better than foreign instrument-RTC158.
Subject
General Physics and Astronomy
Reference10 articles.
1. The influence of axial temperature distribution on calibration accuracy based on dry block furnace;Zhao;11th International Conference on Prognostics and System Health Management,2020
2. Dry block calibrator using heat flux sensors and an adiabatic shield;Hohmann;International Journal of Thermophysics,2011
3. Dry block calibrator with improved temperature field and integrated fixed-point cells;Marin;International Journal of Thermophysics,2012
4. A semi-micro dry-block and automated analyser technique suitable for determining protein nitrogen in plantmaterial;Starr;Journal of Agricultural Science,2008
5. Design, construction, and analysis of a continuous-temperature infrared calibrator for temperature measurement using an infrared scanner;Hsieh;Review of Scientific Instruments,2012