Author:
Trisna Beni Adi,Suherlan ,Wiriadinata Hidayat,Fajria Melati Azizka,Rifa’i Iip Ahmad,Tistomo Arfan Sindhu,Zaid Ghufron
Abstract
Abstract
The thermoelectric inhomogeneity as a function of position along wires is one the significant uncertainty of measurement using thermocouples. Here we report development of an electrical annealing system for thermoelectric inhomogeneity treatment. Two inhomogeneous type-S thermocouples, which had the inhomogeneity greater than 0.04%emf, are successfully recovered using the system. An improvement on thermocouple performance as large as 0.28 °C (at temperature of 1000 °C) can be obtained using the system. This article provides detailed information and may help the reader to obtain a quick grasp about the system.
Subject
General Physics and Astronomy
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献