Abstract
Abstract
We use a continuous neutral density attenuator to
investigate the photomultiplier tube (PMT) charge nonlinearity which
exists in the PMT response to external signals to be measured. The
results show that this testing approach provides a reliable and
practical method for characterizing the PMT nonlinearity, enabling
optimization of the operation conditions and improving the
measurement accuracy of PMT by mitigating the nonlinearity. This
study helps to advance the research and measurements in the related
fields using a PMT.