Author:
Andreotti M.,Capelli S.,Cavallero G.,Chiozzi S.,Cotta Ramusino A.,D'Ambrosio C.,Fiorini M.,Franzoso E.,Frei C.,Gallorini S.,Gambetta S.,Giugliano C.,Gotti C.,Gys T.,Keizer F.,Maino M.,Malecki B.,Minzoni L.,Mitchell S.,Neri I.,Petrolini A.,Piedigrossi D.,Robertson G.,Sergi A.,Simi G.,Slazyk I.,Smith M.,Webster J.,Wotton S.A.
Abstract
Abstract
Signal-induced noise is observed in Hamamatsu R11265
Multianode Photomultiplier Tubes, manifesting up to several
microseconds after the single photoelectron response signal and
localised in specific anodes. The mean number of noise pulses varies
between devices, and shows significant dependence on the applied
high-voltage. The characterisation of this noise and the mitigation
strategies to perform optimal single-photon counting at 40 MHz, as
required by the LHCb Ring-Imaging Cherenkov detectors, are
reported.
Subject
Mathematical Physics,Instrumentation
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. First results on the performance of the upgraded LHCb RICH detectors;Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment;2024-11
2. Systematic approach to measure the performance of microchannel-plate photomultipliers;Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment;2024-08
3. The LHCb Upgrade I;Journal of Instrumentation;2024-05-01
4. Performance of the most recent Microchannel-Plate PMTs for the PANDA DIRC detectors at FAIR;Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment;2023-12
5. Characterisation and operations of the Multianode Photomultiplier Tubes for the LHCb RICH detectors;Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment;2023-11