Shaker Hesham. H.,Kasban H.,Saleh A.A.,Dessouky M.
Abstract
Abstract
Gamma spectrometer is not available in many laboratories
because it is very expensive. Analog to digital converter ADC is the
basic component of digital gamma spectrometers. The sampling rate of
the ADC plays an important role in determining the performance and
the cost of the gamma spectrometers. This paper investigates
experimentally the ADC's sampling rate effect on the performance of
the gamma spectroscopy. The conducted experiment started by sampling
a real pre-amplifier output. Then, the acquired samples have been
processed at different sampling rates to build the dedicated gamma
spectrums. During the processing of the sampled signals, three
different digital filters have been tested to distinguish their
effect on the energy resolution when the sampling rate is
reduced. The results of this experimental investigation will help
later in proposing new low-cost and efficient gamma spectrometer
prototypes.