Author:
Gaganov V.V.,Ryabeva E.V.,Molodtsev D.A.,Ibragimov R.F.,Vershinin I.S.,Kokorev Y.A.,Urupa I.V.
Abstract
Abstract
The results of testing and characterisation of the neutron
spectrometer based on a diamond detector are presented. It is shown
that spectrometer calibration with monoenergetic electrons and
alpha-particles leads to different results. The energy resolution of
the spectrometer is estimated based on the experiments with
alpha-particles and with DT-neutrons from NG-150M neutron
generator. It has been shown experimentally that for neutrons with
an energy of 14 MeV the energy resolution of a diamond-based
spectrometer is 0.6%.
Subject
Mathematical Physics,Instrumentation
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献