Author:
Dai Wei,Tan Mengqing,Zhu Huanan
Abstract
Abstract
Due to the solar radiation effect, the current thermometric
instruments may produce a radiation error of about 1°C. In
this paper, a new radiation shield was designed. Initially, a
Computational Fluid Dynamics (CFD) method was employed to create the
radiation shield structure. Then, the CFD method was used to
quantify the radiation errors of the radiation shield under various
environmental conditions. Next, a Multi-layer Perceptron (MLP)
network was employed to form a radiation error correction approach
for this radiation shield. Finally, the air temperature observation
experiments were performed to verify the accuracy of the new
radiation shield. A 076B fan aspirated radiation shield was served
as an air temperature reference during experiments. The mean
absolute error, root mean square error, and correlation coefficient
between the radiation error results provided by the correction
approach and the measured radiation error results given by the
experiments are 0.012°C, 0.015°C, and 0.796°C,
respectively.
Subject
Mathematical Physics,Instrumentation