Author:
Zhang Xianpeng,Song Zhaohui,Zhang Jianfu,Liu Linyue,Liu Jinliang,Yi Han,Chen Yonghao,Jiang Wei
Abstract
Abstract
The fission target detection system, which is based on
U-235(or U-238)fission target and large area 4H-SiC detector, has
been developed for pulsed neutron detection. However, calibration of
its neutron energy response curve is challenging due to its very low
detection efficiency to fast neutrons. In the past, the calibration
of the neutron sensitivity for the fission target detection system
was limited to using ∼ 14 MeV neutrons generated by the
Cockcroft-Walton accelerator. To obtain the response function for
the complete range of neutron energies, extrapolation based on
neutron fission cross sections was required. In this paper, we
present the calibration of its energy response at the China
Spallation Neutron Source (CSNS) Back-n beam line using the time of
flight method, both with double-bunch and single-bunch mode in
end-station 1(with a flight path of ∼ 55 m). The SiC detector
with a thin sensitive thickness of 30 μm has lower scattering
background than the traditional Si-PIN detector . As a result, the
energy response curve of the fission target detection system above
0.26 MeV has been successfully acquired. The spectrum of the
double-bunch mode was unfolded using the Bayesian algorithm, and the
results indicate that, under conditions of sufficient counts, this
method can yield results similar to those obtained from the
single-bunch mode. The issue of pulse height defect of SiC detector
to fission fragments is discussed. This experiment confirms that our
detection system of SiC+fission target has a strong signal-to-noise
ratio and anti-interference capability. At the same time, it
demonstrates that this device has its own advantages in measuring
the energy response of low sensitivity detectors.
Subject