Towards 2D dosimetry using monolithic active pixel sensors and a copper grating

Author:

Beck L.,Velthuis J.J.,De Sio C.,Pritchard J.L.,Li Y.,Hugtenburg R.P.

Abstract

Abstract Higher energy and intensity X-ray radiotherapy treatments are coming into wider use, having the benefit of requiring fewer treatment fractions and fewer hospital visits per patient. However, small percentage errors in multileaf collimator positioning and dose become bigger problems with higher doses per fraction. Hence, real-time treatment verification becomes essential. Where devices downstream from the patient suffer from scattering in the patient, upstream devices can disturb the therapeutic beam. Here, a method is proposed for performing dosimetry upstream using monolithic active pixel sensors, which can be made thin enough to disturb the beam by <1%. In order to calculate the dose to the tumour, a verification device needs to make a measurement of the photon field. Some photons will Compton scatter an electron in the silicon and generate a signal. However, this signal is obscured by energy deposits from contamination electrons, originating from Compton scattering in the accelerator head and air. Often extensive build-up material is added to verification devices to reduce the electron contamination and enhance the photon signal. However, this leads to degradation of the beam intensity to the patient. Instead we propose using thin strips of 50 µm thick copper in a grating pattern and measuring the difference in the signal with and without it. The contamination electrons are relatively undisturbed by the presence of the thin copper strips and the photon signal generated via Compton scattering is enhanced under the strips. Hence the difference in the two signals mostly consists of energy deposits originating from the therapeutic photons. From this the dose to the patient can be derived. The 50 μm of copper is thin enough to keep the beam attenuation below 1%, but in itself would give a total signal response which consists of 38% contamination electrons. Using the grating technique, we show that the electron contamination can be reduced to 2.6% of the total signal. This allows the photon signal only to be extracted from the data and thus the dose to patient with a very thin upstream detector can be calculated using a Monte Carlo model to extrapolate the photon flux into the tumour.

Publisher

IOP Publishing

Subject

General Medicine

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3