Abstract
Abstract
Higher energy and intensity X-ray radiotherapy treatments are coming into wider use, having the benefit of requiring fewer treatment fractions and fewer hospital visits per patient. However, small percentage errors in multileaf collimator positioning and dose become bigger problems with higher doses per fraction. Hence, real-time treatment verification becomes essential. Where devices downstream from the patient suffer from scattering in the patient, upstream devices can disturb the therapeutic beam. Here, a method is proposed for performing dosimetry upstream using monolithic active pixel sensors, which can be made thin enough to disturb the beam by <1%. In order to calculate the dose to the tumour, a verification device needs to make a measurement of the photon field. Some photons will Compton scatter an electron in the silicon and generate a signal. However, this signal is obscured by energy deposits from contamination electrons, originating from Compton scattering in the accelerator head and air. Often extensive build-up material is added to verification devices to reduce the electron contamination and enhance the photon signal. However, this leads to degradation of the beam intensity to the patient. Instead we propose using thin strips of 50 µm thick copper in a grating pattern and measuring the difference in the signal with and without it. The contamination electrons are relatively undisturbed by the presence of the thin copper strips and the photon signal generated via Compton scattering is enhanced under the strips. Hence the difference in the two signals mostly consists of energy deposits originating from the therapeutic photons. From this the dose to the patient can be derived. The 50 μm of copper is thin enough to keep the beam attenuation below 1%, but in itself would give a total signal response which consists of 38% contamination electrons. Using the grating technique, we show that the electron contamination can be reduced to 2.6% of the total signal. This allows the photon signal only to be extracted from the data and thus the dose to patient with a very thin upstream detector can be calculated using a Monte Carlo model to extrapolate the photon flux into the tumour.
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献