Author:
Hu Q.,Ye H.Q.,Zhang Q.L.,Yi G.H.,Chen L.,Xu X.Y.,Jin G.
Abstract
Abstract
A compact and integrated neutron-gamma discrimination spectrometer that could be utilized
for personal dosimetry and emergency response is urgently required. In this study, based on a CLYC
scintillator coupled to silicon photomultiplier(SiPM) arrays, we designed a stand-alone, hand-held
neutron-gamma discrimination spectrometer. After the analog-to-digital converter (ADC) digitizes
the analog signal from the detector, the data is transmitted to a FPGA processor for online
real-time processing, and then transmitted to a microcontroller through the serial peripheral
interface(SPI) protocol, and the current detection result is displayed in real-time on the touch
screen. By using the pulse signal discrimination(PSD) method, the spectrometer can simultaneously
measure the spectrum of neutron and gamma events, with a resolution of 8.8% (at 1274.5 keV)
and excellent linearity. The figure of merit (FOM) is 2.32 with optimized PSD parameters. The
experimental results show that the spectrometer integrates the upper computer and detector, with
real-time performance, portability and excellent neutron-gamma resolution. Online operation is
realized through the touch screen, which greatly facilitates the real-time detection of radiation
in the field environment.