Author:
Kewlani H.M.,Gharat S.H.,Roychowdhury P.,Dikshit B.,kumar Rajesh
Abstract
Abstract
The ion beam extracted from a proton ion source contains
H+ ion along with H2
+ and H3
+ molecular ions. In this
work, the ionic fractions in the extracted beam of ECR ion source
were measured using an analyzing magnet. A computerized proton
fraction measurement setup is described in detail with
instrumentation and control for pulsed operation of the ECR ion
source. A timing diagram has been incorporated for trigger
synchronization illustrating the process of pulsing the plasma,
controlling the analyzing magnet parameters and acquiring data about
the magnetic field and beam current. The proton fraction was
measured in two cases, in the first case using SS304 plasma chamber
and in the second case incorporating aluminium (Al) cylinder and
boron nitride (BN) plates inside the plasma chamber. In the first
case, the proton fraction was found to be in the range of ∼60%
and in the second case because of the presence of secondary electron
donors it increased to 93%. In order to understand the enhancement
of H+ fraction in the hydrogen plasma, plasma parameters (ne
,
Te
and EEDF) and its dependence on microwave power and neutral
gas pressure were investigated.