Abstract
Abstract
This paper proposes an efficient low cost mechanism to accurately measure the dielectric
dissipation factor and its allied effect over resin impregnated paper. The measuring circuit is a
modified De Sauty bridge network where the accurate parameter measurement can been performed. The
effect of stray capacitance over dielectric loss and dissipation factor has been nullified by the
application of operational amplifier in the circuit. Different samples of resin impregnated paper
have been tested based on their aging and moisture content to check the reliability of the
proposed measuring technique. Experimental result shows the efficacy of the proposed method when
compared to existing IEC and ASTM standard procedures and commercial measuring
devices. Measurement error analysis with repeatability and uncertainty calculation shows promising
result and is within the acceptable limits.
Subject
Mathematical Physics,Instrumentation