Author:
de Souza G.G.A.,Abelha T.S.,Saramela T.B.,Cortez A.F.V.,da Luz H.N.,Penteado C.G.,Bregant M.
Abstract
Abstract
In this work we present the latest developments in the SAMPA-SRS integration. A
software was developed to improve the acquisition configuration, acquisition, and decoding of the data. The complete framework was tested using a triple GEM-based position sensitive detector for X-rays. The detector was operated in Ar/CO2 (70/30) in continuous flow, at atmospheric pressure and made use of a 1 dimension strip readout (200 μm wide strips at a pitch of 400 μm) for charge collection. With this detector a position resolution of better than 833 μm was obtained, with an energy resolution of 14.2 % (σ/E) for 5.9 keV.
Subject
Mathematical Physics,Instrumentation