Author:
Sazzad A.B.M.R.,Busenitz J.,Piepke A.,Poudel S.,Trewin H.,LeViness A.
Abstract
Abstract
In this paper we describe the development, testing, and characterization of three
low-emission rate AmLi neutron sources. The sources are used to calibrate the nuclear recoil
response of the LUX-ZEPLIN (LZ) dark matter experiment. The sources' neutron emission rate was
measured using 3He proportional tubes. The sources' gamma emissions were
characterized using a high-purity germanium (HPGe) detector. Source-validated Geant4
Monte Carlo simulations allowed to calibrate the Ge and neutron detectors' responses.
Subject
Mathematical Physics,Instrumentation
Reference34 articles.
1. The LUX-ZEPLIN (LZ) Experiment;Akerib;Nucl. Instrum. Meth. A,2020
2. LUX-ZEPLIN (LZ) Technical Design Report;Mount,2017
3. Characterization of two extraordinary AmLi neutron sources
4. Reactions Li-7 (alpha, n) B-10, Li-7 (alpha, alpha-prime) Li-7*, and B-10 (n, alpha) Li-7;Bichsel;Phys. Rev.,1957
5. The 7 Li(α,n) 10 B differential cross section for α-energies of up to 8 MeV;van der Zwan;Nucl. Phys. A,1972
Cited by
1 articles.
订阅此论文施引文献