Abstract
Abstract
PandaX-4T is a dark matter direct detection experiment
located in China jinping underground laboratory. The central
apparatus is a dual-phase xenon detector containing 4 ton liquid
xenon in the sensitive volume, with about 500 photomultipliers
instrumented in the top and the bottom of the detector. In this
paper we present a completely new system of readout electronics and
data acquisition in the PandaX-4T experiment. Compared to the one
used in the previous PandaX dark matter experiments, the new system
features triggerless readout and higher bandwidth. With triggerless
readout, dark matter searches are not affected by the efficiency
loss of external triggers. The system records single photelectron
signals of the dominant PMTs with an average efficiency of 96%, and
achieves the bandwidth of more than 450 MB/s. The system has been
used to successfully acquire data during the commissioning runs of
PandaX-4T.