Author:
Garty G.,Harken A.D.,Brenner D.J.
Abstract
Abstract
Standard dosimetry protocols exist for highly penetrating
photon and particle beams used in the clinic and in
research. However, these protocols cannot be directly applied to
shallow penetration MeV-range ion beams. The Radiological Research
Accelerator Facility has been using such beams for almost 50 years
to irradiate cell monolayers, using self-developed dosimetry, based
on tissue equivalent ionization chambers.
To better align with the internationally accepted standards, we
describe implementation of a commercial, NIST-traceable, air-filled
ionization chamber for measurement of absorbed dose to water from
low energy ions, using radiation quality correction factors
calculated using TRS-398 recommendations. The reported dose does not
depend on the ionization density in the range of
10–150 keV/μm.
Subject
Mathematical Physics,Instrumentation
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献