Abstract
Abstract
An innovative SiC Schottky junction and a traditional p-n Si surface barrier detector have been compared to detect carbon ions with MeVs kinetic energy. To this, a comparison was performed during Rutherford backscattering spectrometry (RBS) using 2–10 MeV carbon ion beams. The energy resolution and detection efficiency for RBS analysis using the two detectors and their detection electronics are presented.
The detector parameters dependencies on the surface passivating layers, ion energy and current dependence, ion penetration depth, detection efficiency, energy resolution, and others are discussed.
The comparison of RBS analysis with SiC and Si is investigated highlighting the advantages and disadvantages of using SiC with respect to the traditional Si surface barrier detectors. The two detectors employed for proton, helium and carbon RBS spectrometry of different targets have been also compared on the base of the literature data.