Abstract
Abstract
Radiation measurement relies on pulse detection, which can
be performed using various configurations of high-speed
analog-to-digital converters (ADCs) and field-programmable gate
arrays (FPGAs). For optimal power consumption, design simplicity,
system flexibility, and the availability of DSP slices, we consider
the Radio Frequency System-on-Chip (RFSoC) to be a more suitable
option than traditional setups. To this end, we have developed
custom RFSoC-based electronics and verified its feasibility. The
ADCs on RFSoC exhibit a flat frequency response of 1–125 MHz. The
root-mean-square (RMS) noise level is 2.1 ADC without any digital
signal processing. The digital signal processing improves the RMS
noise level to 0.8 ADC (input equivalent
40 μVrms). Baseline correction via digital
signal processing can effectively prevent photomultiplier overshoot
after a large pulse. Crosstalk between all channels is less than
-55 dB. The measured data transfer speed can support up to
32 kHz trigger rates (corresponding to 750 Mbps). Overall, our
RFSoC-based electronics are highly suitable for pulse detection, and
after some modifications, they will be employed in the Kamioka
Liquid Scintillator Anti-Neutrino Detector (KamLAND).