Author:
Cieślak M.J.,Gamage K.A.A.,Glover R.,Taylor C.J.
Abstract
The pulse shape discrimination performance of a pixelated organic plastic scintillator has been investigated. The scintillator has been built using 169 plastic scintillator blocks (arranged into a 13 × 13 square array) of 2.8 × 2.8 × 15 mm3 each. The scintillator was coupled with a single-channel photomultiplier tube. The scintillator was exposed to a mixed-field environment provided by 252Cf and its pulse shape discrimination capabilities are presented in this paper. Initial results revealed that a 150 MS/s digitising system was insufficient to separate neutrons from gamma-ray photons. Therefore, the experiment was repeated with a 500 MS/s system, which provided improved pulse shape discrimination performance. In order to validate the performance of the pixelated plastic scintillator, it was compared to that of a cylindrical plastic sample. Tests were also carried out in moderated neutron and gamma-ray fields of 252Cf. The results indicate that acceptable levels of pulse shape discrimination are obtained for the case of a pixelated scintillator, when the higher sampling rate digitiser was used.
Subject
Mathematical Physics,Instrumentation
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献