This study is focused on an investigation of the performance
of the Short Strip module developed by the ATLAS Inner Tracker (ITk)
strip collaboration using electron beams of energy 5.4 GeV and
5.8 GeV at the DESY-II Testbeam Facility. The noise at
+30 °C and -30 °C was
measured. The ratio of the two measurements is compared with a
circuit-model calculation. The measured noise at
-30 °C is compared with the maximum noise that
would correspond to an acceptable signal-to-noise ratio after the
expected radiation damage from operation at HL-LHC. The measured
charge distributions at +30 °C and
-30 °C are compared with GEANT4 simulations. The
detection efficiency and noise-occupancy were measured as a function
of threshold at both +30 °C and
-30 °C. The average cluster width was measured
as a function of threshold. Scans of detection efficiency versus
threshold at different delay settings were used to reconstruct the
pulse shape in time. The resulting pulse shape was compared with a
circuit model calculation.