Simulated sensor characterization for virtual clinical trials in mammography and digital breast tomosynthesis

Author:

Sarno A.,Tucciariello R.M.

Abstract

Abstract Virtual clinical trials in X-ray breast imaging permit to compare different technical solutions and imaging modalities at reduced costs, involved personnel, reduced times and reduced radiation risks to patients. In this context, the detector characteristics (spatial resolution, noise level and efficiency) play a key role for an appropriate generation of simulated images. The project AGATA proposes to compute images as dose deposit maps in a detector layer of defined materials. Simulated images are then post-processed on the basis of suitable comparison between intrinsic characteristics of real and simulated detectors. With this scope, as first step for the post-processing manipulations, we evaluated the presampled modulation transfer function (MTF), the detector-response function and the noise power spectrum (NPS) of the simulated detectors. Two detectors were simulated: (1) 0.20 mm-thick a-Se direct flat panel with 70 µm pixel pitch and (2) CsI(Tl) indirect flat panel with 100 µm pixel pitch and scintillator layer 0.25 mm thick. In addition, the impact of simulating the de-excitation processes (Auger emission and fluorescence) was explored. Simulated detector characteristics were evaluated for W/Rh spectra between 25 kV and 31 kV. The in-silico platform used a Monte Carlo software based on Geant4 toolkit (vers. 6). First, the simulation and tracking of electrons generated from photoelectric or Compton interactions was shown to have neglectable influence on the pixel values for the explored spectra, with the produced electrons presenting short ranges with respect to the pixel dimension. In the case of the CsI detector, which has fluorescence energies higher than those of the simulated X-ray photons, the deexcitation processes have not noticeable influence on the calculated pixel values. On the other hand, the MTF of the a-Se detector resulted slightly lower when the fluorescence is simulated in the detector materials, due to the dose spread derived from the fluorescence photons, which can travel far from the initial ionization interaction. Regarding the a-Se detector, the noise power spectrum resulted lower with simulated deexcitation.

Publisher

IOP Publishing

Subject

Mathematical Physics,Instrumentation

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. A Model for a Linear a-Se Detector in Simulated X-Ray Breast Imaging With Monte Carlo Software;IEEE Transactions on Radiation and Plasma Medical Sciences;2024-03

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3