Uniformity of response of Uranium fission chambers used as neutron beam monitors
-
Published:2022-01-01
Issue:01
Volume:17
Page:P01032
-
ISSN:1748-0221
-
Container-title:Journal of Instrumentation
-
language:
-
Short-container-title:J. Inst.
Author:
Lafont F.,Guerard B.,Hall-Wilton R.,Kanaki K.
Abstract
Abstract
Uranium fission chambers are commonly used for neutron beam
monitoring in neutron research facilities. The challenges brought
with the renewing or commissioning of new neutron facilities
encourage to minimize the interaction between the monitor and the
neutron beam. In order to characterize both their detection
efficiency and transmission factor, several fission chambers, have
been scanned on the CT1 monochromatic beam line of the ILL. Results
show that the monitor transparency, measured at the center of the
monitor, varies from 90% to 96%, and that the detection efficiency
varies by as much as 50% over the active area.
Subject
Mathematical Physics,Instrumentation
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献