DH-Mammo PET: a dual-head positron emission mammography system for breast imaging

Author:

Shi YuORCID,Wang Yirong,Zhou Jianwei,Zhu Yuzhen,Lyu Xudong,Hui HuiORCID,Wen Bo,Liu Yanyun,Li Lei,Li Juntao,Meng Fanzhen,Kang Fei,Zhu ShoupingORCID

Abstract

Abstract Objective. To develop a simultaneous positron emission tomography-Optical (OPET) breast imaging dual-head PET subsystem, called DH-Mammo PET, for accurate, early diagnosis and efficacy assessment of breast cancer with high resolution and sensitivity. Approach. We developed a breast-dedicated PET based on LYSO crystal, silicon photomultiplier array and multi-voltage threshold sampling technique. It consists of two detector heads, each with a detection area of 216 mm × 145.5 mm. The distance between the detector heads is fixed at 120 mm. In order to extract coincidences and correct data, GPU-based software coincidence processing, random, scatter, normalization, gap-filling and attenuation corrections were applied in turn. The images were reconstructed using maximum likelihood expectation maximization with depth of interaction (DOI) modeling. The performance of DH-Mammo PET was evaluated referring to NEMA NU 4–2008, NU 2–2007 and Chinese industry recommended standard YY/T 1835–2022. Besides, several clinical patient images of DH-Mammo PET were compared with those of a whole-body PET/CT. Main results. The energy resolution was 14.5%, and time resolution was < 1.31 ns. Indicated by the 22Na point source imaging, its spatial resolution was 2.60 mm (5.40 mm), 1.00 mm (1.04 mm), and 0.96 mm (0.93 mm) in the X, Y and Z directions, respectively, using the system response matrix with (without) DOI modeling. Indicated by the Derenzo phantom imaging, the spatial resolution was ∼3.0 mm, <1.2 mm, and <1.2 mm in the X, Y and Z directions. The system sensitivity was 6.87%, 4.89% and 3.37% with an energy window of 100–800, 250–750 and 350–650 keV, respectively. The scatter fraction was 26.43%, and the peak NECR was 162.6 kcps at 24.1 MBq for the modified rat-like phantom. As for the recovery coefficients, they ranged from 0.15 to 1.04 for rods between 1 and 5 mm obtained with a NEMA image quality phantom. The spill-over ratio for the air-filled and water-filled chamber was 0.05 and 0.11, respectively. DH-Mammo PET can provide more image details in clinical experiments and fulfil a fast scan with 60–120 s acquisition time. Significance. Good spatial resolution and high sensitivity of DH-Mammo PET would enable fast and accurate PET imaging of the breast. Besides, combining the DH-Mammo PET with the diffuse optical tomography would make full use of tumor metabolic imaging and tissue endogenous optical imaging, which would improve the accuracy of early clinical diagnosis of small lesions of breast cancers.

Funder

Fundamental Research Funds for the Central Universities

National Key Research and Development Program of China

CAS Key Technology Talent Program

CAS Youth Innovation Promotion Association

Natural Science Foundation of Hebei Province

Natural Science Foundation of Shaanxi Province

National Natural Science Foundation of China

Publisher

IOP Publishing

Subject

Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology

Reference56 articles.

1. Influence of statistical fluctuation on reproducibility and accuracy of SUVmax and SUVpeak: a phantom study;Akamatsu;J. Nucl. Med. Technol.,2015

2. Effects of noise, image resolution, and ROI definition on the accuracy of standard uptake values: a simulation study;Boellaard;J. Nucl. Med.,2004

3. Initial characterization of a dedicated breast PET/CT scanner during human imaging;Bowen;J. Nucl. Med.,2009

4. Correction methods for random coincidences in fully 3D whole-body PET: impact on data and image quality;Brasse;J. Nucl. Med.,2005

Cited by 5 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3