A method using 4D dose accumulation to quantify the interplay effect in lung stereotactic body radiation therapy

Author:

Huesa-Berral Carlos,Burguete Javier,Moreno-Jiménez Marta,Diego Azcona Juan

Abstract

Abstract The purpose of this study was to devise and evaluate a method to quantify the dosimetric uncertainty produced by the interplay between the movement of multileaf collimator and respiratory motion in lung stereotactic body radiation therapy. The method calculates the dose distribution for all control points from a dynamic treatment in all respiratory phases. The methodology includes some characteristics of a patient’s irregular breathing patterns. It selects, for each control point, the phases with maximum and minimum mean dose over the tumor and their corresponding adjacent phases, whenever necessary. According to this selection, the dose matrices from each control point are summed up to obtain two dose distributions in each phase, which are accumulated in the reference phase subsequently by deformable image registration (DIR). D 95 and D min , 0.035 cc were calculated over those accumulated dose distributions for Gross Tumor Volume (GTV), Planning Target Volume—based on Internal Target Volume approach—and Evaluation Target Volume (ETV), a novel concept that applies to 4D dose accumulation. With the ETV, DIR and interplay uncertainties are separated. The methodology also evaluated how variations in the breathing rate and field size affects the mean dose received by the GTV. The method was applied retrospectively in five patients treated with intensity modulated radiotherapy—minimum area defined by the leaves configuration at any control point was at least 4 cm2. Uncertainties in tumor coverage were small (in most patients, changes on D 95 and D min , 0.035 cc were below 2% for GTV and ETV) but significant over- and under-dosages near ETV, which can be accentuated by highly irregular breathing. Uncertainties in mean dose for GTV tended to decrease exponentially with increasing field size and were reduced by an increase of breathing rate. The implementation of this method would be helpful to assess treatment quality in patients with irregular breathing. Furthermore, it could be used to study interplay uncertainties when small field sizes are used.

Funder

Instituto de Salud Carlos III

Asociación de Amigos de la Universidad de Navarra

Publisher

IOP Publishing

Subject

Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3