Online MR-guided proton and ion beam radiotherapy: investigation of image quality

Author:

Paul KORCID,Dorsch SORCID,Elter AORCID,Beyer C,Naumann J,Hansmann T,Feldmeier E,Haberer T,Karger C P,Debus J,Klüter SORCID

Abstract

Abstract Objective. Magnetic resonance (MR) images free of artefacts are of pivotal importance for MR-guided ion radiotherapy. This study investigates MR image quality for simultaneous irradiation in an experimental setup using phantom imaging as well as in-vivo imaging. Observed artefacts are described within the study and their cause is investigated with the goal to find conclusions and solutions for potential future hybrid devices. Approach. An open MR scanner with a field strength of 0.25 T has been installed in front of an ion beamline. Simultaneous magnetic resonance imaging and irradiation using raster scanning were performed to analyze image quality in dedicated phantoms. Magnetic field measurements were performed to assist the explanation of observed artifacts. In addition, in-vivo images were acquired by operating the magnets for beam scanning without transporting a beam. Main Results. The additional frequency component within the isocenter caused by the fringe field of the horizontal beam scanning magnet correlates with the amplitude and frequency of the scanning magnet steering and can cause ghosting artifacts in the images. These are amplified with high currents and fast operating of the scanning magnet. Applying a real-time capable pulse sequence in-vivo revealed no ghosting artifacts despite a continuously changing current pattern and a clinical treatment plan activation scheme, suggesting that the use of fast imaging is beneficial for the aim of creating high quality in-beam MR images. This result suggests, that the influence of the scanning magnets on the MR acquisition might be of negligible importance and does not need further measures like extensive magnetic shielding of the scanning magnets. Significance. Our study delimited artefacts observed in MR images acquired during simultaneous raster scanning ion beam irradiation. The application of a fast pulse sequence showed no image artefacts and holds the potential that online MR imaging in future hybrid devices might be feasible.

Funder

Bundesministerium für Bildung und Forschung

Publisher

IOP Publishing

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3