Objective. To develop a bremsstrahlung target and megavoltage (MV) x-ray irradiation platform for ultrahigh dose-rate (UHDR) irradiation of small-animals on the Advanced Rare Isotope Laboratory (ARIEL) electron linac (e-linac) at TRIUMF. Approach. An electron-to-photon converter design for UHDR radiotherapy (RT) was centered around optimization of a tantalum–aluminum (Ta–Al) explosion-bonded target. Energy deposition within a homogeneous water-phantom and the target itself were evaluated using EGSnrc and FLUKA MC codes, respectively, for various target thicknesses (0.5–1.5 mm), beam energies (E
e− = 8, 10 MeV) and electron (Gaussian) beam sizes (
2
σ
= 2–10 mm). Depth dose-rates in a 3D-printed mouse phantom were also calculated to infer the compatibility of the 10 MV dose distributions for FLASH-RT in small-animal models. Coupled thermo-mechanical FEA simulations in ANSYS were subsequently used to inform the stress–strain conditions and fatigue life of the target assembly. Main results. Dose-rates of up to 128 Gy s−1 at the phantom surface, or 85 Gy s−1 at 1 cm depth, were obtained for a 1 × 1 cm2 field size, 1 mm thick Ta target and 7.5 cm source-to-surface distance using the FLASH-mode beam (E
e− = 10 MeV, 2
σ
= 5 mm, P = 1 kW); furthermore, removal of the collimation assembly and using a shorter (3.5 cm) SSD afforded dose-rates >600 Gy s−1, albeit at the expense of field conformality. Target temperatures were maintained below the tantalum, aluminum and cooling-water thresholds of 2000 °C, 300 °C and 100 °C, respectively, while the aluminum strain behavior remained everywhere elastic and helped ensure the converter survives its prescribed 5 yr operational lifetime. Significance. Effective design iteration, target cooling and failure mitigation have culminated in a robust target compatible with intensive transient (FLASH) and steady-state (diagnostic) applications. The ARIEL UHDR photon source will facilitate FLASH-RT experiments concerned with sub-second, pulsed or continuous beam irradiations at dose rates in excess of 40 Gy s−1.
Funder
Natural Sciences and Engineering Research Council of Canada
New Frontiers in Research Fund
Canada Research Chairs
Subject
Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology
Reference58 articles.
1. The TRIUMF ARIEL RF modulated thermionic electron source;Ames,2016
2. Introducing the Voxel Interactive Contour Tool for Online Radiation Intensity Analytics (VICTORIA);Badun,2021
3. Modality comparison for small animal radiotherapy: a simulation study;Bazalova;Med. Phys.,2014
Cited by
14 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献