Simulation study on the performance of time-over-threshold based positioning in monolithic PET detectors

Author:

Thyssen CharlotteORCID,Deprez Karel,Mollet Pieter,Van Holen Roel,Vandenberghe Stefaan

Abstract

Abstract The vast majority of PET detectors in the field today are based on pixelated scintillators. Yet, the resolution of this type of detector is limited by the pixel size. To overcome this limitation, one can use monolithic detectors. However, this detector architecture demands specific and high-speed detector readout of the photodetector array. A commonly used approach is to integrate the current pulses generated by every pixel but such circuitry quickly becomes bulky, power consuming and expensive. The objective of this work is to investigate a novel readout and event positioning scheme for monolithic PET detectors, based on time-over-threshold (ToT). In this case, we measure the time that the pulse is above a certain threshold through a comparator. The pulse widths are used for event positioning using a mean nearest neighbour approach (mNN ToT ). For energy determination one integrating multiplexed channel is foreseen. We evaluate the positioning accuracy and uniformity of such a ToT detector by means of Monte Carlo simulations. The impact of the threshold value is investigated and the results are compared to a detector using mean nearest neighbour with pulse-integration (mNN int ), which has already proven to allow sub-mm resolution. We show minimal degradation in spatial resolution and bias performance compared to mNN int . The highest threshold results in the worst resolution performance but degradation remains below 0.1 mm. Bias is largely constant over different thresholds for mNN ToT and close to identical to mNN int . Furthermore we show that ToT performs well in terms of detector uniformity and that scattered photons can be positioned inside the crystal with high accuracy. We conclude from this work that ToT is a valuable alternative to pulse-integration for monolithic PET detectors. This novel approach has an impact on PET detector development since it has the advantage of lower power consumption, compactness and inherent amplitude-to-time conversion.

Funder

Agentschap Innoveren en Ondernemen

Publisher

IOP Publishing

Subject

Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology

Reference33 articles.

1. Understanding and simulating SiPMs;Acerbi;Nuclear Inst. and Methods in Physics Research, A,2019

2. Scintillation camera with multichannel collimators;Anger;J. Nucl. Med.,1964

3. Maximum-likelihood methods for processing signals from gamma-ray detectors;Barrett;IEEE Trans. Nucl. Sci.,2009

4. Experimental validation of an efficient fan-beam calibration procedure for k-nearest neighbor position estimation in monolithic scintillator detectors;Borghi;IEEE Trans. Nucl. Sci.,2015

5. Towards monolithic scintillator based TOF-PET systems: practical methods for detector calibration and operation;Borghi;Physics in Medicine and Biology. IOP Publishing,2016

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Performance and Security Enhancement Solutions for Positron Emission Tomography Medical Hardware;2022 IEEE 16th International Conference on Anti-counterfeiting, Security, and Identification (ASID);2022-12-02

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3