Geometric uncertainty analysis of MLC tracking for lung SABR

Author:

Caillet Vincent,Zwan Benjamin,Briggs Adam,Hardcastle Nicholas,Szymura Kathryn,Prodreka Alexander,O’Brien Ricky,Harris Ben E,Greer Peter,Haddad Carol,Jayamanne Dasantha,Eade Thomas,Booth Jeremy,Keall Paul

Abstract

Abstract Purpose. The purpose of this work was to report on the geometric uncertainty for patients treated with multi-leaf collimator (MLC) tracking for lung SABR to verify the accuracy of the system. Methods. Seventeen patients were treated as part of the MLC tracking for lung SABR clinical trial using electromagnetic beacons implanted around the tumor acting as a surrogate for target motion. Sources of uncertainties evaluated in the study included the surrogate-target positional uncertainty, the beam-surrogate tracking uncertainty, the surrogate localization uncertainty, and the target delineation uncertainty. Probability density functions (PDFs) for each source of uncertainty were constructed for the cohort and each patient. The total PDFs was computed using a convolution approach. The 95% confidence interval (CI) was used to quantify these uncertainties. Results. For the cohort, the surrogate-target positional uncertainty 95% CIs were ±2.5 mm (−2.0/3.0 mm) in left-right (LR), ±3.0 mm (−1.6/4.5 mm) in superior–inferior (SI) and ±2.0 mm (−1.8/2.1 mm) in anterior–posterior (AP). The beam-surrogate tracking uncertainty 95% CIs were ±2.1 mm (−2.1/2.1 mm) in LR, ±2.8 mm (−2.8/2.7 mm) in SI and ±2.1 mm (−2.1/2.0 mm) in AP directions. The surrogate localization uncertainty minimally impacted the total PDF with a width of ±0.6 mm. The target delineation uncertainty distribution 95% CIs were ±5.4 mm. For the total PDF, the 95% CIs were ±5.9 mm (−5.8/6.0 mm) in LR, ±6.7 mm (−5.8/7.5 mm) in SI and ±6.0 mm (−5.5/6.5 mm) in AP. Conclusion. This work reports the geometric uncertainty of MLC tracking for lung SABR by accounting for the main sources of uncertainties that occurred during treatment. The overall geometric uncertainty is within ±6.0 mm in LR and AP directions and ±6.7 mm in SI. The dominant uncertainty was the target delineation uncertainty. This geometric analysis helps put into context the range of uncertainties that may be expected during MLC tracking for lung SABR (ClinicalTrials.gov registration number: NCT02514512).

Funder

National Health and Medical Research Council

Varian Medical Systems

Publisher

IOP Publishing

Subject

Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3