Iterative metal artifact reduction on a clinical photon counting system—technical possibilities and reconstruction selection for optimal results dependent on the metal scenario

Author:

Anhaus Julian AORCID,Schmidt Simon,Killermann Philipp,Mahnken Andreas,Hofmann Christian

Abstract

Abstract Objective. To give an overview about technical possibilities for metal artifact reduction of the first clinical photon-counting CT system and assess optimal reconstruction settings in a phantom study, assessing monoenergetic imaging (VMI) and iterative metal artifact reduction (iMAR). Approach. Scans were performed with 120 kV and Sn140 kV on the first clinical photon-counting detector CT scanner. To quantify artifact reduction, anthropomorphic phantoms (hip, dental, spine, neuro) were assessed, in addition to a tissue characterization phantom (Gammex) to quantify the HU restoration accuracy, all with removable metal inserts. Each setup was reconstructed with and without dedicated iMAR, and VMIs were computed in 10 keV steps from 40 keV (60 keV at Sn140 kV) to 190 keV for all setups (ground truth and metal with and without iMAR). To find the optimal energy, pixel-wise errors were computed in relevant ROIs in water-equivalent tissue around the metal in each phantom setup. To assess HU restoration potential, measurements were performed in the Gammex phantom’s inserts. Main results. Large metal objects (hip head) or metal with high atomic numbers (dental and neuro) do not benefit from higher-energetic reconstructions. The hip shaft (large, low atomic number) comprises a lower base artifact level than the head, still without an energetic optimum. Within the spine (short penetration length, low atomic number) an energy optimum could be identified for both spectra (100 keV for 120 kV and 120 keV for Sn140 kV). The Gammex showed best HU restoration at 100 keV for 120 kV and at 110 keV for Sn140 kV. In all cases, additional iMAR reduced the base artifact level. Significance. This study shows that a novel photon-counting CT system has the capability to reduce metal artifacts in metal types with low atomic number and low penetration length by applying VMI. For all other metal types, additional iMAR is required to reduce artifacts.

Publisher

IOP Publishing

Subject

Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology

Cited by 19 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3