Accuracy and potential improvements of surface-guided breast cancer radiotherapy in deep inspiration breath-hold with daily image-guidance

Author:

Nankali SaberORCID,Hansen Rune,Worm Esben,Yates Esben S,Thomsen Mette Skovhus,Offersen Birgitte,Poulsen Per Rugaard

Abstract

Abstract Objective. Radiotherapy of left-sided breast cancer in deep inspiration breath-hold (DIBH) reduces the heart dose. Surface guided radiotherapy (SGRT) can guide the DIBH, but the accuracy is subject to variations in the chest wall position relative to the patient surface. Approach. In this study, ten left-sided breast cancer patients received DIBH radiotherapy with tangential fields in 15–18 fractions. After initial SGRT setup in free breathing an orthogonal MV/kV image pair was acquired during SGRT-guided breath-hold. The couch was corrected to align the chest wall during another breath-hold, and a new SGRT reference surface was acquired for the gating. The chest wall position error during treatment was determined from continuous cine MV images in the imager direction perpendicular to the cranio-caudal direction. A treatment error budget was made with individual contributions from the online registration of the setup MV image, the difference in breath-hold level between setup imaging and SGRT reference surface acquisition, the SGRT level during treatment, and intra-fraction shifts of the chest wall relative to the SGRT reference surface. In addition to the original setup protocol (Scenario A), SGRT was also simulated with better integration of image-guidance by capturing either the new reference surface (Scenario B) or the SGRT positional signal (Scenario C) simultaneously with the setup MV image, and accounting for the image-guided couch correction by shifting the SGRT reference surface digitally. Main results. In general, the external SGRT signal correlated well with the internal chest wall position error (correlation coefficient >0.7 for 75% of field deliveries), but external-to-internal target position offsets above 2 mm occasionally occurred (13% of fractions). The PTV margin required to account for the treatment error was 3.5 mm (Scenario A), 3.4 mm (B), and 3.1 mm (C). Significance. Further integration of SGRT with image-guidance may improve treatment accuracy and workflow although the current study did not show large accuracy improvements of scenario B and C compared to scenario A.

Publisher

IOP Publishing

Subject

Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3